ClearanceBrief
Recall Z-0955-2008 ยท LKK Pump, Infusion, Implanted, Programmable

Medtronic Neuromodulation

Medtronic SynchroMed EL Programmable Pumps, Model 8627L-10, Medtronic, Inc., 710 Medtronic Parkway, Minneapolis, MN 55432-5604 USA. (The implantable Medtronic SynchroMed EL Programmable Pump is part of the SynchroMed EL Infusion System designed to contain and administer parenteral drugs to a specific site)

DateFeb 9, 2008 (initiated Aug 3, 2007)
StatusTerminated
ReasonPump Motor Stall; pumps can stall due to gear shaft wear.
Root causeDevice Design
ActionBegining August 3, 2007, consignees were notified by letter. The letter "Medtronic Urgent: Medical Device Correction Aug 2007" described the problem and the Affected Devices as well as Patient Risk, Patient Management Recommendations, Next Steps and Physician and Patient Support.
Quantity81,023 total for recalls Z-0950-0957-2008
DistributionWorldwide-USA including states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, DS, TN, TX, UT, VA, VT, WA, WI, WV, and WY, and countries of HONG KONG, JAPAN, AUSTRALIA, CANADA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.