ClearanceBrief
Recall Z-1045-2008 ยท LNH System, Nuclear Magnetic Resonance Imaging

Siemens Medical Solutions USA, Inc

Siemens Magentom Symphony, Model Number: 4760265, Siemens Medical Solutions USA, Malvern PA, 19355

DateMar 25, 2008 (initiated Dec 1, 2007)
StatusTerminated
ReasonMagnet Quench, unintended- magnet of a mobile MRI system quenched into the exam room.
Root causeDevice Design
ActionThe firm issued an Update Instruction package to consignees 12/07. As an immediate measure, the recalling firm is initiating field checks of mobile MRI systems per Update Instruction MR039/07/S. This update defines a safety inspection for mobile systems, including checks for potential ice formation and proper configuration of venting assemblies.
Quantity2 units
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K970852MAGNETOM PROJECT 024 SYSTEM (NUMARIS 3.5 SOFTWARE)
Siemens Medical Solutions USA, Inc., Jun 5, 1997