Recall Z-1154-2008 ยท LKK Pump, Infusion, Implanted, Programmable
Medtronic Neuromodulation
Medtronic IsoMed Implantable Constant-Flow Infusion Pump, model 8472-60. The contents of the inner package have been sterilized by ethylene oxide gas. Reservoir size: 60 mL. Medtronic, Inc., 710 Medtronic Parkway NE, Minneapolis, MN 55432-5604, USA. The implantable Medtronic IsoMed Constant-Flow Infusion Pump is part of the IsoMed Constant-Flow Infusion System designed to contain and admini
| Date | Mar 22, 2008 (initiated Jan 16, 2008) |
| Status | Terminated |
| Reason | Device/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of intrathecal catheters which infuse opioids, baclofen, or chemotherapy drugs into patients. On January 16, 2008, Medtronic sent a letter to doctors who implant these d |
| Root cause | Device Design |
| Action | An Urgent Medial Device Correction letter was sent January 16, 2008, to Health Care Professionals. The letter describes the incidences, symptoms and recommendations for patient management. Excerpts from the approved Medtronic Professional Labeling are also included with the letter. The firm has requested unreported inflammatory mass in a patient with a Medtroinc device to the firm and to the FDA MedWatch Program by phone at 1-800-FDA-1088. Additional assistance may be obtained by contacting Medtronic Neuromodulation Technical Services at 1-800-707-0933. |
| Quantity | 102,792 worldwide. 90,330 within US, 12,462 OUS |
| Distribution | Worldwide Distribution -- USA including states of Washington D.C., Puerto Rico, and countries of Aruba, Australia, Austria, Belarus, Belgium, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Faroe Islands, Finland, France, French Polynesia, Germany, Greece, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.