ClearanceBrief
Recall Z-1146-2008 ยท LKK Pump, Infusion, Implanted, Programmable

Medtronic Neuromodulation

Medtronic SynchroMed EL Implantable Infusion Pump, model 8627-10. The contents of the inner package have been sterilized by ethylene oxide gas. 10 mL pump with reservoir valve and screened side catheter access port. Medtronic, Inc., 710 Medtronic Parkway NE, Minneapolis, MN 55432-5604, USA. The implantable Medtronic SynchroMed EL Programmable Pump is part of the SynchroMed EL Infusion System

DateMar 22, 2008 (initiated Jan 16, 2008)
StatusTerminated
ReasonDevice/Drug Interaction - The company updated the labeling for the devices to include current patient management and treatment recommendations. The company received reports of inflammatory mass formations at or near the distal tip of intrathecal catheters which infuse opioids, baclofen, or chemotherapy drugs into patients. On January 16, 2008, Medtronic sent a letter to doctors who implant these d
Root causeDevice Design
ActionAn Urgent Medial Device Correction letter was sent January 16, 2008, to Health Care Professionals. The letter describes the incidences, symptoms and recommendations for patient management. Excerpts from the approved Medtronic Professional Labeling are also included with the letter. The firm has requested unreported inflammatory mass in a patient with a Medtroinc device to the firm and to the FDA MedWatch Program by phone at 1-800-FDA-1088. Additional assistance may be obtained by contacting Medtronic Neuromodulation Technical Services at 1-800-707-0933.
Quantity~102,792 worldwide. 90,330 within US and 12,462 OUS
DistributionWorldwide Distribution -- USA including states of Washington D.C., Puerto Rico, and countries of Aruba, Australia, Austria, Belarus, Belgium, Brazil, Canada, China, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Egypt, Faroe Islands, Finland, France, French Polynesia, Germany, Greece,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.