Recall Z-0369-2008 ยท JAA System, X-Ray, Fluoroscopic, Image-Intensified

GE Healthcare

GE Precision 500D Classical Radiographic-Fluoroscopic X-ray System. Model #2289299-2, with console computer and touch screen fullly integrated. the GE Precision 500D Classical Radiographic-Fluoroscopic X-ray System is used to perform radiographic and fluoroscopic x-ray examinations.

DateSep 4, 2008 (initiated Oct 4, 2007)
StatusTerminated
ReasonIt was reported that when the GE Precision 500D foot pedal is repeatedly activated, a defect in the fluoro-timer will cause the dose measurement to report a higher value than was actually received by the patient. Patient will receive a lower dosage irradiation time which will result in indication of cumulative air kerma higher than that to which the patient was actually exposed.
Root causeRadiation Control for Health and Safety Act
ActionA customer letter "GE Healthcare Product Safety Notification" dated 10/4/07 was distributed to customers informing them of the safety issue. A Field Modification Instructions (FMI 10806) was issued to the field for the Precision 500D product to install a sofware upgrade to correct the issue. This correction was released in October 2007.
Quantity106
DistributionAL, AR, CA, CO, CT, DC, FL, GA, IA, ID, IL, IN, KY, LA, MA, MI, MN, MS, MT, NC, NE, NJ, NM, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI, ITALY, Kuwait, Great Britan, Spain and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011624EXPEDIO 500D R&F X-RAY SYSTEM
GE Medical Systems, Jun 8, 2001