Recall Z-1188-2008 ยท DQA Oximeter

Kentec Medical Inc

Reusable Ohmeda Compatible Finger Clip Sensor, Model Number: E403-02, Product Code: PS2225. To monitor oxygen saturation level of patients under doctor's care. Kentec Medical Inc., Irvine, CA 92614

DateJun 26, 2008 (initiated May 18, 2007)
StatusTerminated
ReasonIncompatible components: The Finger Clip Sensors - with a Serial Number prefix of PA514, PA515, YE508, ZC502 and ZG512 - are missing an additional diode and higher resistor in its existing configuration making them inoperable when used to interface with those Ohmeda Tuffsat monitors previously re-configured by Ohmeda to include a detection circuit for diodes.
Root causeProcess change control
ActionConsignees were notified via telephone on 5/18/07 and an "Important Voluntary Market Withdrawal Notice" letter on 5/23/07. The letter instructed consignees to segregate and return all affected devices. The letter included an "Acknowledgement of Market Withdrawal Notification" form for customer to complete and return to the recalling firm. For additional information, contact 1-800-825-5996.
Quantity41
DistributionNationwide Distribution including states of AZ, CA, CO, FL, KY, LA, MA, NC, NJ, TX and UT.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K970098E100 SERIES SPO2 SENSORS
Epic Medical Equipment Services, Inc., Apr 30, 1998