Recall Z-1139-2008 ยท LZS Excimer Laser System

VISX INCORPORATED, A SUBSIDIARY OF AMO INC

AMO VISX STAR Excimer Laser System, manufactured by VISX, Santa Clara, CA 95051

DateJun 5, 2008 (initiated Dec 13, 2007)
StatusTerminated
ReasonFirm received two complaints (including one injury complaint) associated with the movement of the VISX chair while in the patient loading ("fully swung out" unlocked) position under an IntraLase FS laser.
Root causeComponent design/selection
ActionOn December 13, 2007, firm issued an Urgent Safety Notice to all VISX and IntraLase laser owners related to recall. The Safety Notice contains information related to the measures that can be taken to prevent unintentional movement of any chair during other procedures. AMO is initiating a field corrective action, whereby Field Service Engineers will visit each practice that owns both systems to disable the VISX chair back switch. For additional information, please contact AMO at (714) 247-8656.
Quantity1,853 units
DistributionWorldwide Distribution - USA, including the states of: AK AL AR AZ CA CO CT DC DE FL GA HI IA ID IL IN KS KY LA MA MD MI MN MO MS MT NC ND NE NH NJ NM NV NY OH OK OR PA PR RI SC SD TN TX UT VA VT WA WI WV and countries of: Argentina Australia Austria Bahrain Belgium Bolivia Brazil Bulgaria Canada Ch
FDA recordOpen on fda.gov