Recall Z-1166-2008 ยท CGZ Electrode, Ion-Specific, Chloride

Abbott Laboratories, Inc

Clinical Chemistry ICT Calibrator, List Number: 1E46-02, Abbott Labs, Abbott Park, IL 60064

DateMay 24, 2008 (initiated Sep 26, 2007)
StatusTerminated
ReasonQuality Control (Out of Range)- ICT Calibrator lot number: 0505017, may contain an interferent. Not all vials within a kit are affected. The investigation determined that the interferent affects Potassium (K+) only. Use of the affected calibrator vials will generate an acceptable, but low calibration slope. When Quality Control (QC) is run to verify the calibration, K+ QC results may be bel
Root causeComponent design/selection
ActionAn Immediate Action Product Recall dated 9/26/07, was provided to all AEROSET and ARCHITECT cSYSTEM customers that have received Lot Number 0505017. The effectiveness check will be completed through a customer reply form for 100% of US customers. Outside of the U.S. each individual country will determine the methods of effectiveness check. Necessary Actions: If the ICT calibration generates an acceptable slope for (K+), and the calibration is verified by generating QC results within the acceptable range, then the customers may continue to use the lot until a new one is available. If the I
Quantity1,377 Distributed Worldwide; 419 Domestically
DistributionWorldwide Distribution --- including USA, Puerto Rico and countries of Mexico, Ecuador, Brazil, Canada, Hong Kong, New Zealand, Venezuela, Chile, , Germany, Thailand, Trinidad/Tobago, Columbia, Argentina, El Salvador, Singapore, Australia, and the Bahama.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K980367AEROSET SYSTEM
Abbott Laboratories, Apr 1, 1998