Recall Z-1244-2008 ยท LLZ System, Image Processing, Radiological
AGFA Corp.
WEB1000 Versions 3.1, 4.1 and 5.1, Picture Archiving and Communications systems (PACs), used in the acceptance, transfer, display, storage and digital processing of medical images.
| Date | Sep 11, 2008 (initiated May 14, 2007) |
| Status | Terminated |
| Reason | Java version 1.6x (JAVA 6) is not compatible with AGFA IMPAX Administration/Service Tools. Due to a Software Mismatch data can be lost. |
| Root cause | Software change control |
| Action | Agfa Healthcare contacted all consignees on 5/14/2007 and by registered mail on 06/07/2007. The registered Urgent Field Safety Notice letters requested the accounts take the following steps immediately: 1. Turn off automatic JAVA updates to avoid this update being applied in the background. 2. Do not manually update JAVA on their station-even for other applications. 3. If JAVA 1.6 has already been applied, do not use Service/Admin Tools. They were instructed to contact their IT department or Agfa support for assistance in backing out of this version to apply the required version for the Agfa s |
| Quantity | 946 units (total) |
| Distribution | U.S. Nationwide and Canada. |
| FDA record | Open on fda.gov |