ClearanceBrief
Recall Z-1244-2008 ยท LLZ System, Image Processing, Radiological

AGFA Corp.

WEB1000 Versions 3.1, 4.1 and 5.1, Picture Archiving and Communications systems (PACs), used in the acceptance, transfer, display, storage and digital processing of medical images.

DateSep 11, 2008 (initiated May 14, 2007)
StatusTerminated
ReasonJava version 1.6x (JAVA 6) is not compatible with AGFA IMPAX Administration/Service Tools. Due to a Software Mismatch data can be lost.
Root causeSoftware change control
ActionAgfa Healthcare contacted all consignees on 5/14/2007 and by registered mail on 06/07/2007. The registered Urgent Field Safety Notice letters requested the accounts take the following steps immediately: 1. Turn off automatic JAVA updates to avoid this update being applied in the background. 2. Do not manually update JAVA on their station-even for other applications. 3. If JAVA 1.6 has already been applied, do not use Service/Admin Tools. They were instructed to contact their IT department or Agfa support for assistance in backing out of this version to apply the required version for the Agfa s
Quantity946 units (total)
DistributionU.S. Nationwide and Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K022292IMPAX DIAGNOSTIC DISPLAY STATION,IMPAX CLINICAL REVIEW STATION,IMPAX FOR CARDIOLOGY CLINICAL REVIEW SATATION,IMPAX ORTHO
Agfa Corp., Sep 12, 2002
K040555IMPAX CLIENT EMBRACE
Agfa Corp., May 26, 2004
K050228AGFA HEARTLAB CARDIOVASCULAR
Heartlab, Inc., Feb 18, 2005
K050751IMPAX ORTHOPAEDIC WORKSTATION, MODEL IMPAX ES 5.X
Agfa Corp., Apr 21, 2005
K053458WEB1000
Agfa Corp., Dec 22, 2005