Recall Z-1243-2008 ยท LLZ System, Image Processing, Radiological

AGFA Corp.

IMPAX 4.1, IMPAX 4.5, IMPAX 5.0, IMPAX 5.1 IMPAX 5.2, IMPAX 6.0, IMPAX 6.2; Picture Archiving and Communications systems (PACs), used in the acceptance, transfer, display, storage and digital processing of medical images.

DateSep 11, 2008 (initiated May 14, 2007)
StatusTerminated
ReasonJava version 1.6x (JAVA 6) is not compatible with AGFA IMPAX Administration/Service Tools. Due to a Software Mismatch data can be lost.
Root causeSoftware change control
ActionAgfa Healthcare contacted all consignees on 5/14/2007 and by registered mail on 06/07/2007. The registered Urgent Field Safety Notice letters requested the accounts take the following steps immediately: 1. Turn off automatic JAVA updates to avoid this update being applied in the background. 2. Do not manually update JAVA on their station-even for other applications. 3. If JAVA 1.6 has already been applied, do not use Service/Admin Tools. They were instructed to contact their IT department or Agfa support for assistance in backing out of this version to apply the required version for the Agfa s
Quantity946 units (total for both types of software)
DistributionU.S. Nationwide and Canada.
FDA recordOpen on fda.gov