ClearanceBrief
Recall Z-0884-2008 ยท DQX Wire, Guide, Catheter

Argon Medical Devices, Inc

Argon Medical Devices, ArgoGuide Hydrophilic NiTi Guidewire, K011905, Catalog Number: 00-AG387308, .035/180cm, stiff/straight/standard, sold as a single hydrophilic guidewire in both sterile and bulk non-sterile packaging configurations. Argon Medical Devices, Inc. Athens, TX 75751

DateJan 31, 2008 (initiated Nov 19, 2007)
StatusTerminated
ReasonDegradation -- Hydrophilic guidewires may exhibit degradation of blue Pebax cladding causing particulate to dislodge potentially resulting in embolism.
Root causeProcess design
ActionArgon Medical sent an Urgent Product Recall letter to consignees on 11/15/2007. They requested that users immediately quarantine and return any product remaining at their facility, utilizing instruction on response card. Distributors were directed to issue the recall to their customers and return all retrieved product to Argon.
Quantity121
DistributionWorldwide Distribution -- USA including states of FL, KS, NV, IL, CA, NY, CO, VA, MD, NC, UT, GA, MI, NJ, TN, MA, and CT, and countries of The Netherlands, Germany, and Turkey.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011905SNIPER ELITE MODEL-OR-E3872XX SERIES
Maxxim Medical, Aug 28, 2001