ClearanceBrief
Recall Z-0905-2008 ยท EZN Dilator, Catheter, Ureteral

Boston Scientific Corporation

Pathway Balloon Expandable Ureteral Access Sheath- 11/14/28, Ureteral dilator, Catalog Number: 255100; Product Number: M0062551000, Onset Medical, Boston Scientific Corp., Natick, MA 01760

DateApr 30, 2008 (initiated Nov 30, 2007)
StatusTerminated
ReasonSheath Removal Difficulty: Users experience difficulty in removing the sheath.
Root causeDevice Design
ActionConsignees were sent an Urgent Medical Device Recall letter on November 29, 2007. The letter instructed users to check their inventory, remove any affected product and place them in quarantine. It also included instructions for returning the product. Users were asked to return a Reply Verification Tracking Form. Follow-up communication will be made to users who do not return the tracking form.
Quantity89
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K061009PATHWAY BALLOON EXPANDABLE URETERAL ACCESS SHEATH, MODEL BEUS
Onset Medical Corporation, May 2, 2006