Recall Z-0732-2008 ยท LLZ System, Image Processing, Radiological

Siemens Medical Solutions USA, Inc

Siemens Syngo Imaging System with VB30A version of software. Model Number: 10014063, Siemans Medical Solutions, Malvern, PA 19355

DateFeb 23, 2008 (initiated Nov 27, 2007)
StatusTerminated
ReasonInconsistent data - A buffer overflow may occur between information system and picture system if the system experiences a high load. (e.g. RIS/PACS communication is unusally high)
Root causeSoftware design
ActionThe recalling firm issued a Customer Safety Advisory, via certified mail. to the affected customers on 11/27/07. The letter informs the customer of the issue and also provides instruction to detect and avoid its occurrence. In addition, the firm is developing a software update to correct the issue, which will be installed on the affected systems when it is released.
Quantity10 units
DistributionNationwide Distribution, including states of AL, MO, NC, NY, OH, OR, TN, and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071114SYNGO IMAGING (VERSION V30)
Siemens AG Medical Solutions, May 22, 2007