Recall Z-0311-2009 ยท DTQ Console, Heart-Lung Machine, Cardiopulmonary Bypass

Terumo Cardiovascular Systems Corporation

Terumo Sarns Level Sensor II Pads. For use with Sarns Advanced Perfusion System I, Perfusion System 800 and Perfusion System 900. Terumo Cardiovascular Systems Corp., Ann Arbor, MI.; REF 195240. Level Sensor pads are used to attach the level sensor to the sides of a rigid reservoir.

DateDec 10, 2008 (initiated Jan 13, 2008)
StatusTerminated
ReasonThe level sensor holder may detach from the adhesive pad causing the sensor to lose contact with the venous reservoir, which may result in a detached sensor indication or false alarm.
Root causeNonconforming Material/Component
ActionConsignees were notified via recall letter (Urgent Medical Device Removal letter) dated 2/13/08 and requested to report stocks on hand so that replacements could be sent.
Quantity31,980 pads (25,620 to U.S. and 6,360 international)
DistributionWorldwide Distribution including USA and countries of Australia, Belgium, Canada, China, Dominican Republic, Japan, Saudi Arabia, Singapore and Yemen.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.