Recall Z-1352-2008 ยท MCB Antigen, C. Difficile

Biosite Inc

Triage Micro ClostrIdium difficile (C.difficile) Panel, Catalog No: 96000, BIOSITE INCORPORATED, San Diego, CA 92121

DateMar 27, 2008 (initiated Dec 4, 2007)
StatusTerminated
ReasonFalse Lower Results: Toxin A results in the Triage Micro C. difficile Panel have shown falsely lower visual signals (color strength) in certain lots compared to results observed with previous lots.
Root causeProcess design
ActionConsignees located in the United States received an Urgent Medical Device Recall Letter notification by United Parcel Service on 12/4/2007 and 12/14/2007. Consignees outside of the United States were notified by Federal Express on 12/4/2007 and 12/14/2007. The Customer Recall Letter with attached fax-back Verification form informed customers of the recall and instructed customers to dicontinue use, discard appropriately all affected kits and complete and FAX the enclosed Verification form. Contact information was also provided form the recalling firm. (USA-(800) 640-0640 )
Quantity4,549 kits
DistributionWorldwide- USA, Argentina, Australia, Belgium, Canada, Switzerland, Colombia, The Czech Republic, Germany, France, Greece, Guatemala, Hong Kong, Croatia, India, Italy, Mexico, Portugal, Sweden, Turkey & The United Kingdom
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K974881TRIAGE C. DIFFICILE, TRIAGE C. DIFFICILE PANEL, TRIAGE C. DIFFICILE TOXIN A
Biosite Incorporated, Mar 20, 1998