Recall Z-0710-2008 ยท DQK Computer, Diagnostic, Programmable
McKesson Provider Technologies
Horizon Cardiology Hemo Monitoring System;Horizon Cardiology Hemo External Control Unit;~100 - 230 VAC 50-50 Hz, Medcon Telemedicine Technology LTD.,Tel Aviv, 697190, Israel.
| Date | Feb 5, 2008 (initiated Nov 12, 2007) |
| Status | Terminated |
| Reason | Power supply reliability: McKesson received word from their manufacturer of the medical grade power supply that some power grade supplies may not provide reliable power output. |
| Root cause | Nonconforming Material/Component |
| Action | Recalled by letter sent to the firm's customers beginning on 11-12-2007, advising them that the field correction would be completed by the recalling firm service technicians. |
| Quantity | 13 |
| Distribution | Nationwide to hospital and medical centers in LA. CA, PA and IL, |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K050561 | WINDSURFER Medcon , Ltd., May 4, 2005 |