Recall Z-0668-2008 ยท JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

Stryker Howmedica Osteonics Corp.

Stryker GMRS Proximal Femoral Surgical Protocol; Global Modular Replacement System Proximal Femoral Resection for Large Segmental Replacements, Femur Prosthesis Surgical Literature, Stryker Orthopaedics, Mahwah, NJ 07430

DateApr 22, 2008 (initiated Dec 18, 2007)
StatusTerminated
ReasonThe GMRS Proximal Femoral Surgical Protocol states " The GMRS Proximal Femoral components are fully compatabile with all Stryker V40 femoral heads." This statement is incorrect and conflicts with the Alumina Head label and packaging insert.
Root causeLabeling design
ActionStryker sent Urgent Product Recall notification letters and Product Accountability forms via FedEx on 12/21/2007, return receipt requested. An Urgent Product Recall letter dated 1/7/2008, was sent to surgeons who are GMRS users, advising that the GMRS Proximal femoral components are not compatible with Alumina ceramic heads.
Quantity5869
DistributionWorldwide Distribution - USA, Canada, Mexico, China, Australia, Hong Kong, Singapore, Sweden, Germany, Brussels, the Netherlands, Italy, Poland, Dubai, United Kingdom, Spain, and South Africa.