Recall Z-0875-2008 ยท KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)

Rockwell Medical Technologies, Inc

Dri-Sate Glacial Acidic Acid (liquid component of Dri-Sate Dry Acid Concentrate DR-240); 1 quart (0.95 liters); manufactured by Rockwell Medical Technologies, Inc., Grapevine, TX 76051.

DateFeb 13, 2008 (initiated Dec 3, 2007)
StatusTerminated
ReasonFoil Seal Deterioration; occuring on the tamper resistant foil seals on dialysis liquid concentrate and Dri-Sate product containers older than two years old.
Root causeProcess control
ActionFirm began notification of consignees by letter, Alert: Voluntary Recall, on 12/03/07, sent via e-mail, fax, and/or mail. Consignees asked to check inventory and to identify and call for the replacement of any product older than 2 years old.
Quantity10 cases (1 one quart bottle per case)
DistributionNationwide Distribution to the following states: AL, AR, FL, KS, LA, MO, MS, OK, TN and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K981003ROCKWELL MEDICAL SUPPLY, LLC. DRI-SATE ACID CONCENTRATE MIXES
International Medical Products, Inc., Jun 16, 1998