Recall Z-0873-2008 ยท KPO Dialysate Concentrate For Hemodialysis (Liquid Or Powder)
Rockwell Medical Technologies, Inc
RenalPure Liquid Acid Concentrate R-259; One Gallon (3.78 liters); manufactured by Rockwell Medical Technologies, Inc., Grapevine, TX 76051.
| Date | Feb 13, 2008 (initiated Dec 3, 2007) |
| Status | Terminated |
| Reason | Foil Seal Deterioration; occuring on the tamper resistant foil seals on dialysis liquid concentrate and Dri-Sate product containers older than two years old. |
| Root cause | Process control |
| Action | Firm began notification of consignees by letter, Alert: Voluntary Recall, on 12/03/07, sent via e-mail, fax, and/or mail. Consignees asked to check inventory and to identify and call for the replacement of any product older than 2 years old. |
| Quantity | 6 cases (4 one gallon containers/case). |
| Distribution | Nationwide Distribution to the following states: AL, AR, FL, KS, LA, MO, MS, OK, TN and TX. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K954527 | ROCKWELL MEDICAL SUPPLY LLC HEMODIALYSIS CONCENTRATE SOLUTIONS & POWDERS Rockwell Medical Supply, Mar 1, 1996 |