Recall Z-0893-2008 ยท DYB Introducer, Catheter

Cook, Inc.

Peel-Away Introducer set, 7.0 FR, Catheter introducer, includes sheath: 7.0 FR., 15.5 CM long, Wire: .035" DIA., 50 CM long, 3 MM J needles: 18 ga., 7.0 cm long. Accepts devices up to 7.0 French, sterile, Catalog Number: C-PLI-7.0-35-X-1. Cook Inc., Bloomington, IN 47404

DateApr 16, 2008 (initiated Dec 7, 2007)
StatusTerminated
ReasonPackaging Problem: The peel-away sheath does not peel uniformly or completely.
Root causePackage design/selection
ActionCook Inc. initially notified consignees via FEDEX by an Urgent: Product Recall on 12/7/07. The recall was then expanded and consignees were notified on 12/18/08 by another Urgent: Product Recall. The letters instructed any distributors to inform their customers of recall, and all users to return the recalled products and enclosed Recall Response Form to the firm. For additional information, contact 1-800-346-2686.
Quantity1,881 of total products
DistributionWorldwide Distribution- including USA and countries of Argentina, Australia, Canada, Chile, Denmark, Japan, New Zealand and Singapore.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.