ClearanceBrief
Recall Z-0673-2008 ยท DPW Flowmeter, Blood, Cardiovascular

Parks Medical Electronics Inc

Parks Medical Dual Frequency Doppler Vascular Flow Detector, Model 915-BL (finished assembly without accompanying probes), Parks Medical Electronics, Inc., Aloha, OR 97007

DateJul 3, 2008 (initiated Nov 8, 2007)
StatusTerminated
ReasonPremature battery failure: Wrong acid flux used on circuit board which may lead to premature battery failure of ultrasonic non-directional doppler flowmeter.
Root causeProcess control
Action811-B: On 9/28/07 Parks Medical Electronics notified their related U.S. distributor, Parks Medical Electronics Sales, Inc., Las Vegas, NV by email instructing them to stop further shipments. On 10/1/07 firm followed up with phone call instructing all units be returned for rework. On 11/8/07 firm sent replacement units accompanied with a Warranty Exchange Letter to their other U.S. customers notifying them of affected units. Letters were customized for each customer with a fill-in-the-blank of the model and serial numbers. Letters and replacement units were sent UPS 2nd Day Air and informed
Quantity4 units distributed to 1 U.S. hospital
DistributionWorldwide Distribution - USA, Brazil, Australia, and Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.