Recall Z-1135-2008 ยท FRW Scale, Patient

Arjo, Inc.

BHM Medical Scales, 800 lb. scale weight limit; Part Number 700.00510, Manufactured by BHM Medical Inc., Magog, QC, Canada J1X 5Y5; These scales were sold for installation on or as a component of the following patient lifts: a) Manufactured by BHM Medical Inc., Magog, QC, Canada J1X 5Y5: Part Number: 88570101, Ergolift 600 BAL. Wide Opening Patient Lift; Part Number : ERGOLIFT-600+SC, Ergolift 6

DateFeb 26, 2008 (initiated Dec 19, 2007)
StatusTerminated
ReasonUnintended Hanger Bar Detachment: The locknut on the scales may loosen during use and allow the hanger bar assembly of the patient lift to detach and fall.
Root causeDevice Design
ActionArjo sent Urgent Device Field Correction Notification letters dated 12/19/07 to the end user accounts who received the affected scales, advising them of the potential for the hanger bar assembly to detach. The accounts were given the option immediately stopping use of the lifts with the affected scales or removing the scales from the lift so the lift can continue to be used. Instructions for removal of the scales were provided with the letter. The accounts were requested to complete and return the enclosed customer response form, acknowledging the receipt and understanding of the letter, an
Quantity772 scales
DistributionNationwide including the states of Alabama, Arizona, California, Delaware, Florida, Hawaii, Iowa, Illinois, Indiana, Massachusetts, Michigan, Minnesota, Mississippi, North Carolina, North Dakota, New Hampshire, New Jersey, New York, Ohio, Pennsylvania, South Dakota, Tennessee, Washington and Wiscon
FDA recordOpen on fda.gov