Recall Z-0981-2008 ยท IYE Accelerator, Linear, Medical
Siemens Medical Solutions USA, Inc
Siemens Primeview 3i 2.1 medical charged-particle radiation therapy system, Part No: 8147675, with software version 3i. Siemens AG, Concord, CA.
| Date | Apr 30, 2008 (initiated Oct 2, 2007) |
| Status | Terminated |
| Reason | Image Orientation Incorrect (Image Shifts) - 1) if the reference image has been calibrated for centering and the reference image is used for patient positioning, the image will shift when the positioning tools are used. 2) This problem occurs only when non-square reference images are used, such as when images are cropped. At this point, the images will refresh, and the center of the reference im |
| Root cause | Software design |
| Action | On 10/03/2007, the firm sent Consignees a Customer Safety Advisory Notice informing them of the two potential issues with the display of reference images used for position correction. Consignees will be informed when a solution to this issue has been validated. |
| Quantity | 3 units |
| Distribution | Worldwide Distribution-USA including 48 consignees in the US and 39 outside the US. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K060226 | ONCOR EXPRESSION Siemens Medical Solutions USA, Inc., Mar 15, 2006 |