ClearanceBrief
Recall Z-0977-2008 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

Siemens Syngo based WS 3rd Party V&R medical charged-particle radiation therapy system, Part No: 7345437. Siemens AG, Concord, CA.

DateApr 30, 2008 (initiated Oct 2, 2007)
StatusTerminated
ReasonImage Orientation Incorrect (Image Shifts) - 1) if the reference image has been calibrated for centering and the reference image is used for patient positioning, the image will shift when the positioning tools are used. 2) This problem occurs only when non-square reference images are used, such as when images are cropped. At this point, the images will refresh, and the center of the reference im
Root causeSoftware design
ActionOn 10/03/2007, the firm sent Consignees a Customer Safety Advisory Notice informing them of the two potential issues with the display of reference images used for position correction. Consignees will be informed when a solution to this issue has been validated.
Quantity1 unit
DistributionWorldwide Distribution-USA including 48 consignees in the US and 39 outside the US.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K060226ONCOR EXPRESSION
Siemens Medical Solutions USA, Inc., Mar 15, 2006