ClearanceBrief
Recall Z-0859-2008 ยท FQH Lavage, Jet

Zimmer Orthopaedic Surgical Products

Pulsavac Fan Spray Kit ,Catalog #00-5150-475-01; 1kit/package, Zimmer Orthopaedic Surgical Products, Dover, Ohio

DateMar 6, 2008 (initiated Dec 18, 2007)
StatusTerminated
ReasonSterility (package integrity) compromised; The effect of a silicone stain produced during assembly operations on the sterility barrier properties of the Tyvek lids has not been validated by the firm.
Root causeOther
ActionZimmer notified their customers via an Urgent Device Recall letter dated January 8, 2008. The letter requests each customers to inventory their stock. If any recalled product is located the customer is to notify Zimmer via fax and a replacement part will be shipped to them.
Quantity9,411 kits
DistributionWorldwide-USA, forty-two international and one Canadian account, including countries of Singapore, Hong Kong, Lebanon, Finland, Australia, Japan, Israel, Korea, Saudi Arabia, Dubai, Denmark, Czech Republic, Guatemala, Brazil, Taiwan, China, Nicaragua, Virgin Island, Switzerland, Austria, Spain, Bel
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.