ClearanceBrief
Recall Z-0707-2008 ยท JWH Prosthesis, Knee, Patellofemorotibial, Semi-Constrained, Cemented, Polymer/Metal/Polymer

Biomet, Inc.

Biomet CC Cruciate Tibial Tray, 75 mm.; Product No. 141234, Biomet, Inc., Warsaw, IN 46581

DateJan 31, 2008 (initiated Dec 12, 2007)
StatusTerminated
ReasonMislabeled as to size: Package labeled as containing a 67 mm tray actually contains a 75 mm tray and vice versa.
Root causePackaging process control
ActionConsignees were notified via an Urgent Medical Device Recall Notice dated 12/12/07 to cease use of the product and to return it.
Quantity8
DistributionWorldwide: USA including states of Georgia, Idaho, Indiana, Missouri, Ohio, and Virginia and country of Turkey.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K915132MCK MAXIMUM CONGRUENT KNEE) KNEE SYSTEM
Biomet, Inc., Feb 11, 1992