ClearanceBrief
Recall Z-0680-2008 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

Siemens Mevatron KDS-2 Medical Charged-Particle Radiation Therapy System; (Interlock #51 Jaw Field Position) Model Number 9822693 and 9411588, Siemens Medical Solutions, Concord, CA 94520

DateFeb 14, 2008 (initiated Aug 7, 2007)
StatusTerminated
ReasonIncorrect field size -- Products manufactured prior to 1999 may experience jaw field size errors during interlock resulting in potential mistreatment if a light field check is not performed. This problem is present in machines with serial numbers less than 3094, as they do not contain redundant sensors for jaw position.
Root causeDevice Design
ActionThe firm has sent consignees a Customer Safety Advisory Notice on August 7, 2007 informing them that a misplacement of the X and/or Y jaws can occur if interlock #51 Jaw Field Position is asserted and then cleared with the left-hand key rest. The firm recommends that if an interlock issue is encountered tha treatment of the patient be discontined until this issue is resolved.
Quantity35 units
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K862339MEVATRON KD-2
Siemens Medical Laboratories, Inc., Oct 29, 1986