Recall Z-0592-2008 ยท KMI Monitor, Bed Patient

Hoana Medical

Hoana LifeBed Patient Vigilance System, (Intelligent Medical Vigilance System),Model Number: 010135-01, Product distributed by Hoana Medical, Inc., 828 Fort St. Mall, Suite 620, Honolulu, HI

DateJan 24, 2008 (initiated Nov 7, 2007)
StatusTerminated
ReasonBed Exit Failure: Failure to recognize "Bed Exit" when the feature is active due to a software anomaly.
Root causeSoftware design
ActionOn 11/7/07, all consignees were notified via an untitled letter and on-site visit by the firm's sales representative, informing them of the affected product and providing instructions on the recall. The recalling firm will arrange for installation of a software fix at each consignee.
Quantity216 units
DistributionNationwide including states of HI, CA, TX and MA