Recall Z-0624-2009 ยท IZL System, X-Ray, Mobile

Medtronic Navigation, Inc

Medtronic O-Arm Intraoperative Imaging System Catalog Number: BI-700-00027

DateFeb 4, 2009 (initiated Nov 27, 2007)
StatusTerminated
ReasonNavigation accuracy problem: Accuracy may fall outside of specification when used in combination with the Medtronic Synergy Experience StealthStation System Spine Software with the O-Arm Intraoperative Imaging System and tracker.
Root causeSoftware design
ActionMedtronic Navigation notified Field representatives on 11/27/07 to perform site visits to implement FCA 2007-005 - O-Arm Intraoperative Imaging System Accuracy, and deliver the Customer Letter notification. Confirmation of the visit was documented.
Quantity54 units
DistributionWorldwide Distribution: USA, Canada, Czech Republic, Denmark, Germany, Netherlands, S. Korea, Poland, South Africa, and Singapore
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K050996O-ARM IMAGING SYSTEM
Breakaway Imaging, LLC, May 5, 2005