Recall Z-1099-2008 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use
Beckman Coulter Inc
UniCel DxC 600 PRO Part Number: A10400, Beckman Coulter, Inc.
| Date | May 24, 2008 (initiated Apr 5, 2007) |
| Status | Terminated |
| Reason | Intermittent Failure of Stirrer Motor-Beckman Coulter has determined that incorrect Glucose results on LX20 or DxC instruments have been related to an intermittent failure of the stirrer motor. -Stirrer motors can stall without any flags or motion errors. -If a stall occurs during a GLUm test near the time of sample inject, results may be affected. -Affected GLUm results may be high or low. |
| Root cause | Software design |
| Action | GLUm stir motors greater than one year old were replaced. Replacements began April 5, 2007. Based on the Risk Assessment, BCI decided that Customer Notification is not needed. GLUm stir motors greater than one year old will be replaced on high volume LX/DxC systems by 3rd quarter of 2007. Replacements began April 5, 2007. A new stirrer design is being developed and is expected to be available before 4th quarter of 2007. Implementation Strategy of Stock Control: 1. A Stop Ship was not necessary based on risk assessment. The possibility of stirrer motor failure may increase with time and duty |
| Quantity | 463 in the US; 33 in Canada |
| Distribution | Worldwide Distribution - USA including Canada. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K042291 | UNICEL DXC 600 AND 800 SYNCHRON SYSTEMS Beckman Coulter, Inc., Nov 12, 2004 |