Recall Z-0567-2008 ยท DSP System, Balloon, Intra-Aortic And Control

Datascope Corp

Datascope CS 100 Intra-Aortic Balloon Pump and Datascope CS 300 Intra-Aortic Balloon Pump, Datascope Corp. Mahwah, NJ 07430

DateJan 31, 2008 (initiated Nov 19, 2007)
StatusTerminated
ReasonDefective Circuit Board -- A defect in the Printed Circuit Board's (PCB) in the CS 100 and CS300 Intra-Aortic Balloon Pump may cause the pump to exhibit intermittent malfunctions, which include failure to start-up or reset of the display screen during therapy.
Root causeOther
ActionDatascope sent a Product Field Correction notification letters to Hospital Administrators, via certified mail, on November 19, 2007. The letter informed them of a defect in the circuit board. It described the possible malfunction and provided instructions should malfunctioning occur. Datascope will be replacing the defective circuit boards in all affected IABP units. A Service Representative will be contacting the hospital to arrange for the replacement of the defective board.
Quantity70 units US
DistributionWorldwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063525DATASCOPE FIBER OPTIC PRESSURE SENSOR INTRA-AORTIC BALLOON CATHETER AND CS300 INTRA-AORTIC BALLOON PUMP CONTROL SYSTEM
Datascope Corp., Jan 5, 2007