Recall Z-1367-2008 ยท KDI Dialyzer, High Permeability With Or Without Sealed Dialysate System

Gambro Renal Products, Inc.

Gambro cartridge blood tubing set used with Gambro Dasco Phoenix and Cobe Centrysystem 3 (and 3+) Hemodialysis System, Gambro, Lakewood, CO 80401

DateMar 25, 2008 (initiated Sep 10, 2007)
StatusTerminated
ReasonKinked tubing on hemodialysis device may cause hemolysis.
Root causeDevice Design
ActionOn September 10, 2007, the firm sent an Advisory Notice to all consignees alerting them that a kink in the tubing can result in physical hemolysis. The firm included with the Notice a Quick Reference Guide that provides drawing of the proper installation of the Phoenix and describes how to prevent bloodline kinking. A customer reply form was also included. The firm is in the process of making labeling changes that will be released with the updated software revisions.
Quantity41,000,000
DistributionNationwide, including 17 VA facilities and 1 military facility. Foreign distribution to Canada, Korea, Mexico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K001156PHOENIX HEMODIALYSIS DELIVERY SYSTEM, GAMBRO CARTRIDGE FOR HEMODIALYSIS BLOOD TUBING SET
Gambro Renal Products, Jul 7, 2000