ClearanceBrief
Recall Z-0926-2008 ยท JDI Prosthesis, Hip, Semi-Constrained, Metal/Polymer, Cemented

Stryker Howmedica Osteonics Corp.

Stryker Modular Replacement System Curved Cemented Stem; MRS 15x127 Curved Femoral Stem. Catalog Number: 6485-3-715; Stryker Orthopaedics Mahwah, NJ 07430

DateFeb 26, 2008 (initiated Nov 15, 2007)
StatusTerminated
ReasonCurvature Discrepancy -- The radius of curvature may be out of specification to varying degrees causing a discrepency in the stem curvature between trials and implants.
Root causeProcess design
ActionOn November 26, 2007, Stryker Sent Urgent Product Recall notification letters to Stryker Agencies/Branches and hospitals (OR Supervisor) by Federal Express. The notification requested that all inventory be examined and returned. Returning instructions were provided in the letter. Questions concerning this recall are being addressed by contacting the Regulatory Compliance Hotline at (201) 831-6633.
Quantity803 total, all sizes
DistributionNationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K952970HOWMEDICA FEMORAL & PROXIMAL TIBIAL MODULAR REPLACEMENT SYTESM
Howmedica Corp., Jan 17, 1996