Recall Z-0958-2008 ยท GKZ Counter, Differential Cell

Beckman Coulter Inc

Cytomics FC 500 MPL Flow Cytometry System with MXP Software, Software Version 2.1, Automated differential cell counter, Part Number: 626554, Beckman Coulter, Fullerton, CA 92834

DateApr 30, 2008 (initiated Jun 29, 2007)
StatusTerminated
ReasonData generation error: If the cytosettings is not refreshed or restarted during data importation, the output will include old and new data.
Root causeSoftware design
ActionConsignees were notified by an Urgent Product Corrective Action letter sent on 6/29/07. The letter provided details on the data error and instruction on how to avoid the error. Users were asked to retain notification as part of the Quality Systems documentation and return a response form acknowledging receipt. For additional information, contact 1-800-526-7694.
Quantity516 units in the US; 73 units in Canada
DistributionWorldwide Distribution- USA and country of Canada.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K071681CYTOMICS FC 500 MPL FLOW CYTOMETER, MODEL # 626554
Beckman Coulter, Inc., Oct 4, 2007