Recall Z-0525-2008 ยท GKZ Counter, Differential Cell

Beckman Coulter Inc

Beckman Coulter Flow-Count Fluorospheres - Part Number 7547053, Beckman Coulter, Inc., Fullerton, CA 92835

DateFeb 6, 2008 (initiated May 21, 2007)
StatusTerminated
ReasonSporadic absolute count recovery failures with assayed control cell products when using Lot Number 754801F of the Flow-Count Fluorospheres. Additionally, this lot has experienced secondary fluorescent populations containing more than 20% of the total population, as stated in the Evidence of Deterioration section of the Package Insert.
Root causeProcess design
ActionA Product Corrective Action (PCA) letter was mailed on May 21, 2007, to all accounts that have ordered Flow-Count Fluorospheres, Part Number 7547053, Lot # 754801F. The letter advised that Beckman Coulter has confirmed absolute count recovery failures with assayed control cell products. Additionally, internal testing has indicated secondary fluorescent populations slightly above their indicated 20% of the total population as stated in the Evidence of Deterioration section of the Package Insert. The letter requests that users immediately discard any bottles of the product and call for a repla
Quantity97 units in the US; 37 units Internationally
DistributionWorldwide Distribution --- including USA and country of Canada. ***CENTER COMMENTS -- The Center requests that the subject recall not be terminated until CDRH has determined that the firm has identified the root cause and performed an adequate CAPA to address the cause of the problem, i.e., quality
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K954688FLOW-COUNT COULTER FLUOROSPHERES
Coulter Corp., Feb 29, 1996