Recall Z-0849-2008 ยท HAW Neurological Stereotaxic Instrument

Medtronic Image Guided Neurologics, Inc.

Nexframe Stereotactic System Kits, Model Number(s): DB-1021-MR, DB-1031, DB-1040-BL, DB-1040-ST, DB-1041, DB-1041-BL, DB-1041-ST, DB-1042, DB-1042-BL, DB-1042-ST, DB-1043, DB-1043-BL, DB-1043-ST, DB-2031, DB-2040-BL, DB-2040-ST, DB-2041, DB-2041-BL, DB-2041-ST, DB-2042, DB-2042-BL, DB-2042-ST, DB-2043, DB-2043-BL, DB-2043-ST,, Medtronic, Inc, Minneapolis, Minnesota

DateFeb 7, 2008 (initiated Oct 4, 2007)
StatusTerminated
ReasonSterilty (package integrity) Compromised: Some failures were for damage to the outer pouch, while another set of failures were for the seals on this pouch. The seal between the inner tray and lid has not been compromised and the contents remain sterile, however, the sterility of the outer surfaces of the inner tray and lid cannot be assured.
Root causePackage design/selection
ActionOn 10/4/2007, consignees were notified by Medtronic field personnel, either by personal visits or telephone. All wholesalers of this product are OUS. Medtronic OUS Regulatory personnel communicated this issue directly to the wholesalers, who provided the letters to their customers. Medtronic field personnel are responsible for completing and returning reply cards confirming all notifications (100%).
Quantity385
DistributionWorldwide, USA, Taiwan, Thailand, Japan, Spain, Sweden and The Netherlands.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K012366NAVIGUS II TRAJECTORY GUIDE
Image-Guided Neurologics, Inc., Oct 1, 2001
K022100NAVIGUS MANUAL DISPOSABLE MICRODRIVE WITH DIGITAL DISPLAY, MODELS MI-1000 & MI-1010
Image-Guided Neurologics, Inc., Sep 25, 2002
K040334NAVIGUS PASSIVE PROBE, MODEL NP-1000
Image-Guided Neurologics, Inc., Oct 4, 2004
K043091NAVIGUS NEXDRIVE, MODEL MI-3000
Image-Guided Neurologics, Inc., Feb 7, 2005