Recall Z-0688-2008 ยท JFJ Catalytic Methods, Amylase

Roche Diagnostics Corp.

Roche LIPC Lipase Colorimetric Reagent; Catalog Number: 03029590322. Roche Diagnostics Corp., Indianapolis, IN 46250

DateFeb 8, 2008 (initiated Dec 3, 2007)
StatusTerminated
ReasonFalse Results- The reagent of the LDL-Cholesterol assay shows a carryover effect on the Lipase assay. Results may be falsely elevated by up to 50 units per liter when used on COBAS INTEGRA 400 or COBAS INTEGRA 400+ analyzers.
Root causeProcess design
ActionConsignees were sent and Urgent Medical Device Correction notice dated 12/03/07. The notice provided instructions for conducting an additional wash cycle whenever both the LDL-Cholesterol assays and Lipases assay are used on the COBAS INTEGRA 400 or COBAS INTEGRA 400+ analyzers. Contact Roche Technical Support at 1-800-428-2336 for questions regarding these instructions.
Quantity9,817 kits.
DistributionNationwide Distribution