Recall Z-0382-2008 ยท IYE Accelerator, Linear, Medical
Siemens Medical Solutions USA, Inc
Siemens, Oncor Impression, Part No.: 58 57 920, Medical Charge particle radiation therapy system,with pre-R2.0 Optivue Flat Panel Positioners, Siemens Medical, Concord, CA 94520
| Date | Dec 1, 2007 (initiated Mar 23, 2007) |
| Status | Terminated |
| Reason | Product's collision detection system may not be sensitive enough to trigger a stop action without a significant amount of force. |
| Root cause | Device Design |
| Action | The firm has issued a Customer Safety Advisory Notice dated 12/6/2006, to all sites that have the pre-R2.0 OPTIVUE Flat Panel Positioners installed. Users will have a new microswitch-based collision detection system, installed free of charge. |
| Quantity | 21 units |
| Distribution | Worldwide Distribution- USA including MA, ME, NJ, NY, PA, MD, WV, NC, SC, GA, FL, AL, TN, KY, OH, IN, MI, IA, WI, SD, ND, MT, IL, NE, LA, OK, TX, WY, NV, CA, AK AND AR, and countries of Serbia, Australia, Austria, Belgium, Canada, Denmark, El Salvador, France, Germany, Hungary, India, Ireland, It |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.