Recall Z-0382-2008 ยท IYE Accelerator, Linear, Medical

Siemens Medical Solutions USA, Inc

Siemens, Oncor Impression, Part No.: 58 57 920, Medical Charge particle radiation therapy system,with pre-R2.0 Optivue Flat Panel Positioners, Siemens Medical, Concord, CA 94520

DateDec 1, 2007 (initiated Mar 23, 2007)
StatusTerminated
ReasonProduct's collision detection system may not be sensitive enough to trigger a stop action without a significant amount of force.
Root causeDevice Design
ActionThe firm has issued a Customer Safety Advisory Notice dated 12/6/2006, to all sites that have the pre-R2.0 OPTIVUE Flat Panel Positioners installed. Users will have a new microswitch-based collision detection system, installed free of charge.
Quantity21 units
DistributionWorldwide Distribution- USA including MA, ME, NJ, NY, PA, MD, WV, NC, SC, GA, FL, AL, TN, KY, OH, IN, MI, IA, WI, SD, ND, MT, IL, NE, LA, OK, TX, WY, NV, CA, AK AND AR, and countries of Serbia, Australia, Austria, Belgium, Canada, Denmark, El Salvador, France, Germany, Hungary, India, Ireland, It
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.