Recall Z-0575-2008 ยท NVI Diagnostic Software, K-Nearest Neighbor Algorithm, Autoimmune Disease

Bio-Rad Laboratories Inc

BioRad BioPlex 2200 Software, for BioPlex 2200 Instrument, an automated immunodiagnostic analyzer, Bio-Rad Laboratories, Inc., Hercules, CA 94547

DateApr 22, 2008 (initiated Nov 13, 2007)
StatusTerminated
ReasonIncorrect values: An error was found in the BioPlex 2200 Software, Version 2.0 that results in assignment values being utilized for the BioPlex assay Calibrator Set that may be slightly different from those printed on the Value Assignment Data Sheet. Some values may be identical between the lot assignment combinations with the ANA Screen with MDSS kit.
Root causeSoftware design
ActionBio-Rad sent a Voluntary Field Correction notice dated November 1, 2007. Consignees were notified by phone call on Nov. 13, 2007, followed up by facsimile (domestic). Sub-recall to be by e-mail message for consignees outside the US. Distribution of the product ceased on Nov. 2, 2007. Bio-Rad will provide Calibrator Set CDs with only one lot of calibrator assignment values until a new version of software is available.
Quantity85 units of calibrator lot 86280 and 16 units of calibrator lot 86290 were distributed.
DistributionWorldwide Distribution - USA, Canada, France, and Germany.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043341BIOPLEX 2200 ANA SCREEN WITH MEDICAL DECISION SUPPORT SOFTWARE FOR USE WITH BIOPLEX 2200 MULTI-ANALYTE DETECTION SYSTEM
Bio-Rad Laboratories, Inc., Oct 27, 2005