Recall Z-1510-2008 ยท KPS System, Tomography, Computed, Emission

Philips Medical Systems (Cleveland) Inc

Axis Gamma Camera, Scintillation (gamma) camera, Philips Medical System, Cleveland, OH 44143

DateSep 19, 2008 (initiated Nov 21, 2007)
StatusTerminated
ReasonResistor Failure: The Rotate Motion Shunt Resistor may overheat and results in the appearance of smoke, a burning smell, and an electronic stop condition, which will disable the gamma camera rotate motion and all other motions.
Root causeComponent design/selection
ActionConsignees were sent an Urgent Field Safety Notice on 1/30/08. The letter instructs users who experience a failure to remove the system from service and contact the recalling firm for immediate correction. The recalling firm will schedule appointments with users to replace the affected part beginning in March 2008. For additional information, contact 1-800-722-9377.
QuantityUnknow at this time since the recall was expanded.
DistributionWorldwide Distribution --- USA including states of AL, AZ, CA, CO, CT, DC, DE, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, TN, TX, UT, VA, VT, WA, WI, and WV, Puerto Rico and countries of Argentinia, Austrailia, Austria, Belgiu