Recall Z-0553-2008 ยท KEY Stainer, Tissue, Automated

Ventana Medical Systems Inc

Symphony Staining System (High Volume Stainer), KPA, Model Number 2200000, Ventana Medical Systems Inc

DateJan 25, 2008 (initiated Oct 30, 2007)
StatusTerminated
ReasonFluid Leaks-The T fitting located in the Symphony Clear system of the Symphony High Volume Slide Stainer needs to be larger to securely hold the tubing in place to eliminate any possible fluid leak due to the tubing movement.
Root causeComponent design/selection
ActionThe end user customers were verbally contacted starting in November 2007 to set up a scheduled visit to replace the fitting.
Quantity55
DistributionWorldwide-USA, France and Japan