Recall Z-1159-2008 ยท LKJ Antinuclear Antibody, Antigen, Control

Bio-Rad Laboratories

BIOPLEX 2200 MULTI-ANALYTE DETECTION SYSTEM: a) 100 test - Catalog No. 665-1150; and b) the 5,000 tests - Catalog No. 665-1155 - manufactured by Bio-Rad Laboratories in Redmond, WA. The product is used as an aid in the diagnosis of systemic autoimmune diseases.

DateJun 11, 2008 (initiated Nov 13, 2007)
StatusTerminated
ReasonFalse negative results due to reagent packs exhibiting low signal.
Root causeNonconforming Material/Component
ActionOn 11/13/2007, the firm faxed Voluntary Field Correction Letters (which are dated 11/9/2007) to its consignees and also contacted them via telephone. The Voluntary Field Correction Letters: a) informed them of the potential for false negative results; b) recommended that they perform QC testing daily with each reagent pack; c) provided a new insert for the Instructions for Use; and d) provided a Response Sheet for each consignee to complete and return to the firm.
Quantity8,804 kits
DistributionWorldwide Distribution -- USA, Canada, France, United Kingdom, Italy, Germany, South Africa, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K043341BIOPLEX 2200 ANA SCREEN WITH MEDICAL DECISION SUPPORT SOFTWARE FOR USE WITH BIOPLEX 2200 MULTI-ANALYTE DETECTION SYSTEM
Bio-Rad Laboratories, Inc., Oct 27, 2005