Recall Z-1156-2008 ยท LIP Enzyme Linked Immunoabsorption Assay, Treponema Pallidum

Bio-Rad Laboratories

BIOPLEX 2200 SYPHILIS IgG Kit: 100 tests - Catalog No. 665-1450 - manufactured by Bio-Rad Laboratories in Redmond, WA 98052. The product is a multiplex flow immunoassay intended for the qualitative detection of Treponema pallidium IgG antibodies in human serum.

DateJun 11, 2008 (initiated Nov 13, 2007)
StatusTerminated
ReasonFalse negative results due to reagent packs exhibiting low signal.
Root causeNonconforming Material/Component
ActionOn 11/13/2007, the firm faxed Voluntary Field Correction Letters (which are dated 11/9/2007) to its consignees and also contacted them via telephone. The Voluntary Field Correction Letters: a) informed them of the potential for false negative results; b) recommended that they perform QC testing daily with each reagent pack; c) provided a new insert for the Instructions for Use; and d) provided a Response Sheet for each consignee to complete and return to the firm.
Quantity796 kits
DistributionWorldwide Distribution -- USA, Canada, France, United Kingdom, Italy, Germany, South Africa, and Switzerland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K063866BIOPLEX 2200 SYPHILIS IGG KIT ON THE BIOPLEX 2200 MULTI-ANALYTE DETECTION SYSTEM
Bio-Rad Laboratories, Mar 19, 2007