Recall Z-0551-2008 ยท JQP Calculator/Data Processing Module, For Clinical Use

Abbott Laboratories, Inc

Abbott, ACCELERATOR Device Manager (ADM); List Number: 08H74-01; Abbott Laboratories, Abbott Park, IL

DateJan 25, 2008 (initiated Oct 25, 2007)
StatusTerminated
ReasonSample/Patient Mis-identification: Two issues identified related to the ACCELERATOR Decision Manager (ADM), to include: 1) Sample Identification (SID) and/or Patient Identification (PID) numbers that contain more than 12 characters are truncated to the final 12 characters by the ADM; 2) for SIDs using alpha characters that are case sensitive, SIDs with lower case characters are transmitted by the
Root causeSoftware design
ActionFirm notified consignees via Product Correction, Immediate Action Required letter(s) on 10/25/07. Consignees were notified of software issues and informed they would be contacted by an Abbott representative to schedule installation of software upgrade.
Quantity19 units.
DistributionNationwide:Distributed to hospitals/laboratories in IL, PA, TN, TX, UT, VA and WV.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.