Recall Z-0115-2008 ยท LXH Orthopedic Manual Surgical Instrument

Biomet, Inc.

Biomet Straight Magnum Inserter Handle, a preparatory instrument used for acetabular component insertion; REF # or Catalog Number: S313141. (systems affected by this action include M2a-Magnum, Regenerex, Acetabular Revision Shell and M2a-38), Biomet Orthopedics, Inc., Warsaw, IN. 46581

DateJan 23, 2008 (initiated Sep 4, 2007)
StatusTerminated
ReasonBinding: The instrument was manufactured incorrectly and its use may result in binding to the acetabular component and cause the implant to not release following impaction.
Root causeOther
ActionConsignees were notified by Urgent Medical Device Correction Notice (letter) dated 9/4/07 to discontinue use of the instruments and to return them for evaluation.
Quantity22
DistributionNationwide; including states of Alabama, Arizona, Arkansas, Georgia, Illinois, Indiana, Kentucky, Massachusetts, Michigan, Nevada, North Carolina, Ohio and South Dakota.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.