Recall Z-0484-2008 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Stryker Endoscopy

SERFAS Energy Suction Probe, Model 279-351-100, 3.5mm 90-S, Manufactured by Stryker Puerto Rico

DateJan 17, 2008 (initiated Apr 26, 2007)
StatusTerminated
ReasonDevice Shut off/on: Device may not turn off after taking finger/foot off of activating switch and may activate without pushing a button. (Patient burns may result from these device malfunctions)
Root causeMixed-up of materials/components
ActionConsignees were mailed an Urgent Device Removal Recall notice, dated April 27, 2007. International subsidiaries will receive the recall notice and they will distribute to their consignees. Product is to be returned as a recall.
Quantity6,037 units
DistributionWorldwide Distribution: USA including states of AL, AR, AZ, CA, CO, CT, DE, DC, FL, GA, HI, IA, ID, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, OH, OK, OR, PA, SC, SD, TN, TX, VA, WA, WI, WY, and WV; INTERNATIONAL: Affiliated Stryker Distribution Companies: Sweden (for Scandinav
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K041810SERFAS ENERGY SYSTEM
Stryker Endoscopy, Dec 21, 2004