Recall Z-0520-2008 ยท LLZ System, Image Processing, Radiological

Vital Images, Inc.

Vital Images ViTALConnect , Version 4.0 and 4.1, PET/CT Visualization option. Vital Images, Inc., 5850 Opus Parkway, Suite 300, Minnetonka, MN 55343. ( A web-based medical diagnostic tool that allows physicians to use PCs or notebook computers to gain remote access to 2D, 3D, and 4D advanced visualization. The software enables users to measure, rotate, and analyze images)

DateJan 31, 2008 (initiated Nov 8, 2007)
StatusTerminated
ReasonStandard Uptake Values (SUV) display/update problems; in the way Standard Uptake Values (SUV) are displayed in 3D/MPR views. The problem arises when a user draws a selected region of interest in order to calculate the SUV statistics of the region. If the user then changes the slice shown by scrolling through additional slices, the original SUV measurements will remain on the screen and do not up
Root causeProcess control
ActionCustomers were notified via telephone (11/7 -11/8/2007) and issued an Urgent Medical Device recall notification letter on 11/8/07. The letter described the problem and product involved. The letter also informed customers they they will provide a free software patch within the next 6-8 weeks.
Quantity18
DistributionWorldwide-USA including states of SC, MO, MI, WI, NY, FL, CO, CA, PA, and WA, and countries of England and The Netherlands
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062154VITAL CONNECT 4.0
Vital Images, Inc., Aug 10, 2006
K071362VITAL CONNECT 4.1
Vital Images, Inc., May 30, 2007