Recall Z-0292-2008 ยท DXT Injector And Syringe, Angiographic

Medrad Inc

Medrad Avanta Multi-Patient Disposable Hand Controller, Angiographic Injector and Syringe. Catalog number AVA 500 HC, For Use with MEDRAD Avant Fluid Management System, MEDRAD, Inc., Indianola, PA 15051

DateDec 11, 2007 (initiated Oct 25, 2007)
StatusTerminated
ReasonSterility (package integrity) compromised: Incomplete package seal; sterility may be compromised
Root causePackaging process control
ActionThe recalling firm issued an Urgent Medical Device Field Correction Letter dated 10/25/07 via Federal Express to their customers informing them of the problem. Customers can either return the product or consistent with the indicated use and established pratice of sheathing, utilize sterile sheaths with the hand controllers for up to five patients. Each letter is being accompanied by a box of 25 sheaths.
Quantity25, 977 units
DistributionWorldwide-USA and countries of Germany, Israel, Finalnd, Italy, Netherland, Austria, France, Japan, Denmark, Ireland, South Africia, Australia, Saudi Arabia, Singapore, Canada, New Zealand, Spain, Sweden, Portugal, Korea, Great Britain, Turkey and The United Arab Emerites.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.