ClearanceBrief
Recall Z-0581-2008 ยท KNC Table, Obstetric (And Accessories)

Hill-Rom, Inc.

Hill-Rom Affinity Birthing Bed, Model P3600.

DateFeb 22, 2008 (initiated Nov 21, 2007)
StatusTerminated
ReasonHole/opening on the labor bar -- Amputation of a finger could result, if a finger is placed into the opening, as the hole/opening creates a shear point when the mechanism is activated.
Root causeDevice Design
ActionConsignees were notified of the problem via an Urgent Medical Device Correction letter dated 11/21/07. Users were informed that they would be receiving a free upgrade kit for all of their Affinity Birthing Beds. The Kit includes two plugs per bed, an insertion tool, and full written instructions. An instruction video is available at www.Hill-Rom.com.
Quantity32,379 for all model numbers
DistributionWorldwide Distribution-- including USA and countries of Argentina, Australia, Austria, Bahamas, Bahrain, Belgium, Bermuda, Brazil, Brunei, Canada, Chile, China, Colombia, Costa Rica, Croatia, Cyprus, Czech Republic, Dominican Republic, Egypt, Finland, France, Germany, Greece, Guam, Honduras, Hong Ko
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K915779AFFINITY BED MODEL 3600 SERIES
Hill-Rom, Inc., Feb 24, 1993