ClearanceBrief
Recall Z-0378-2008 ยท JJY Multi-Analyte Controls, All Kinds (Assayed)

Roche Diagnostics Corp.

Roche Ammonia/Ethanol/CO2 Control N (Normal) for use in conjunction with Roche/Hitachi, COBAS INTEGRA, cobas c111 and cobas c 501 clinical chemistry analyzers; Catalog. No. 20752401190. Roche Diagnostics Corporation, Indianapolis, IN 46250

DateDec 29, 2007 (initiated Nov 14, 2007)
StatusTerminated
ReasonHigh Quality Control Recovery: Undetected high quality control recovery in bicarbonate assay.
Root causeProcess design
ActionConsignees were notified of this recall by an Urgent Medical Device Recall letter sent on 11/14/07. The letter instructed the users to immediately discontinue use of Ammonia/ Ethanol/CO2 Control N and Ammonia/Ethanol/CO2 Control A (with listed lots) for quality control assessment of the bicarbonate assay. They were informed that controls may still be used to assess ammonia and ethanol, but alternative controls must be used for assessment of the bicarbonate assay. Consignees are instructed to destroy products at their facility. Consignees were notified of the availability of replacement produ
Quantity811
DistributionNationwide Distibution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K031880AMMONIA/ETHANOL/CO2 CALIBRATOR, CONTROL NORMAL, CONTROL ABNORMAL
Roche Diagnostics Corp., Jul 16, 2003